Last Current(Last_C) Sets the final current value
after OCP has been
tripped. This is the steady-
state current draw after the
OCP has been tripped.
Step Time(Step_T)
Sets the execution time of
each step. (50mS to 1600S)
Delay Time(Delay)
The OCP testing delay
time. Sets the how long to
delay starting the test after
the Load On key has been
pressed. (0 ~ 160S)
Trig Voltage(Trig_V) Sets the voltage trigger
level needed see whether
the power supply OCP has
been triggered.
When the power supply
OCP has been triggered, its
voltage output will drop.
The voltage trigger level is
used to test to see if the
voltage output has been
drop.
Keep Time(Keep_T) Set the how long to enter
the Last Current after
detect the OCP.(0~160S)
Note
This mode can only be used under CC mode.
Panel operation
1.
Press the FUNC key, F4 (OCP) to
enter the OCP Test Automation
menu.
FUNC
F4
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...