OPERATION
227
O
Interface
LOAD
GPIB
GPIB
05/04/22
16 : 50
Address
01
Configure
Other
Interface
System
Info
Load
Time Set
7.
Use the selector knob to edit the
GPIB address.
ENTER
or
8.
Edit the GPIB address.
Range
1 ~ 30
9.
Connect the GPIB cable to the
rear panel port: 24-pin female
connector.
GPIB constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
Note
The GPIB Address must match that of the host
machine.
For GPIB function check, please refer to the section
“GPIB function check” on page 231.
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...