OPERATING DESCRIPTION
89
External Voltage Control
Voltage/
Current
Setting
(Percent
Rating)
External Voltage
100%
0%
10V
0V
To determine the Percent Rating (voltage or
current load input), use the following formula;
VorA
Rating
V
Voltage
External
Input
Load
10
Where “Rating V or A” is the rating
voltage/current of the load module.
Current Monitor
The load current input can be externally monitored
using the IMON pin of a channel control
connector. The IMON pin outputs a voltage of
0~10V to represent the input current as a
percentage (0~100%) of rating current.
Current Monitor
Current
Monitor
Output
Load Input Current
10V
0V
Rating current
0A
To determine the Current Monitor Output
(IMON) , use the following formula;
V
A
Rating
current
input
Load
IMON
10
Where “Rating A” is the rating current of the load
module.
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...