APS-1102A User Manual
APS-1102A
3-26
3.4.6 Setting output frequency
The AC output frequency can be set in the following range.
Table3-11. Output Frequency Setting
Setting Range
Resolution
Setting Accuracy
1.0 Hz to 550.0 Hz
0.1 Hz
0.01% (1.0 Hz to 550.0 Hz, 23
5°C)
S
ee “3.4.7 Using limiter functions”
, when using an output frequency setting range limit.
-------
Notes
--------------------------------------------------------------------------------------------------------
If the frequency is set 40 Hz or lower in AC-INT mode, there is a possibility that the output
waveform will be distorted and DC offset will be output.
----------------------------------------------------------------------------------
Operation steps
To set the output frequency, set
in the SET menu.
1.
Press the
or
key to move the cursor to the
icon.
ENTER
CANCEL
Cursor
2.
Press the
EN TER
key to display the modification box.
ENTER
CANCEL
Modification box