6
REMOTE INTERFACE
APS-1102A
6-45
l) Common commands
CLS
This clears the following registers.
Standard event register
Operation event register
Warning event register
Status byte
Error queue
Syntax
:
CLS
Parameters:
None
Command example:
CLS
Response format:
None
Response example:
None
ESE
This sets the standard event enable register. The setting ranges from 0 to 255.
Syntax:
ESE
<NR1>
Parameters:
0 to 255
Command example:
ESE 255
Query example:
*ESE?
Response format:
<NR1>
Response example:
255
ESR?
This returns the standard event register value. When *ESR reads the standard event register, all
bits in the event register are cleared to 0. The response ranges from 0 to 255.
Syntax:
ESR?
Parameters:
None
Query example:
ESR?
Response format:
<NR1>
Response example:
128
Содержание APS-1102A
Страница 15: ...Tables APS 1102A xiii ...
Страница 16: ......
Страница 24: ......
Страница 34: ...APS 1102A User Manual APS 1102A 2 10 2 5 Calibration To calibrate the APS 1102A contact GW Instek or our agent ...
Страница 72: ...APS 1102A User Manual APS 1102A 3 38 ...
Страница 184: ...APS 1102A User Manual APS 1102A 5 50 ...
Страница 242: ...APS 1102A User Manual APS 1102A 6 58 ...
Страница 266: ...APS 1102A User Manual APS 1102A 8 8 ...