APS-1102A User Manual
APS-1102A
5-26
5.4.3 Output power measurement
The output power values described below are always displayed.
a) Effective power
Full scale:
1200 W
Display resolution:
1 W
Measurement accuracy: at 45 Hz to 65 Hz
(2% of rdg +1 W)
at DC
(3% of rdg +12 W)
(at 23
5°C, load with the power factor 0.5 to lower than 1, 50 V or higher
output voltage, output current is 10% to 100% of the maximum current)
b) Apparent power
Calculated as output voltage RMS value × output current RMS value
Full scale:
1400 VA
Display resolution:
1 VA
Measurement accuracy: at 45 Hz to 65 Hz
(2% of rdg +1 VA)
at DC
(3% of rdg +12 VA)
(at 23
5°C, 50 V or higher output voltage, output current is 10% to 100%
of the maximum current)
c) Reactive power
Calculated as
(
)
apparent power 2-
(
)
effective power 2
Full scale:
1400 var
Display resolution:
1 var
Measurement accuracy:
(2% of rdg +1 var)
(at 23
5°C, load with the power factor 0 to lower than 0.5, 45 Hz to 65 Hz,
50 V or higher output voltage, output current is 10% to 100% of the
maximum current)
5.4.4 Load power factor measurement
Calculated by effective power/apparent power
In the case that it exceeds the measurement range, “
” is displayed.
Measurement range:
0.00 to 1.00
Display resolution:
0.01
Содержание APS-1102A
Страница 15: ...Tables APS 1102A xiii ...
Страница 16: ......
Страница 24: ......
Страница 34: ...APS 1102A User Manual APS 1102A 2 10 2 5 Calibration To calibrate the APS 1102A contact GW Instek or our agent ...
Страница 72: ...APS 1102A User Manual APS 1102A 3 38 ...
Страница 184: ...APS 1102A User Manual APS 1102A 5 50 ...
Страница 242: ...APS 1102A User Manual APS 1102A 6 58 ...
Страница 266: ...APS 1102A User Manual APS 1102A 8 8 ...