5.3.8 RMS current limiter settings
This limits the output RMS current.
When the output RMS current exceeds this setting, the output voltage is suppressed.
Table5-17. RMS Current Limiter Setting Range
Setting
Setting Range
Resolution
Initial Value
RMS current
limiter
100 V range
1.0 A to 10.5 A
0.1 A
10.5 A
200 V range
1.0 A to 5.3 A
0.1 A
5.3 A
Operation steps
Select
in the SET menu and set numerical values.
See
“3.4.7 Using limiter functions”
, for description of the output average value current
limiter.