Designed by:
Drawing by:
Material:
TM
Drawing no.:
Edition:
Sheet:
Surface treatment:
Colour:
Date:
Scale:
Product:
Part:
Tolerance:
fr
937050 wire set
0
29-11-2006
GH2
wire set
A4
Change no.:
92904001
lock washer DIN 6798 A 4mm elzink BN781
1
4
97345410
screw, maskinskrue M4x10 torx
1
3
971254
washer M4 A2
1
2
937049
wire 5000mm with ringshoe
1
1
Drawing/Item no.
Part
Quantity
Pos no.
1
3
2
4
©
Guldmann
GB
-1394/11/08
4.
e
le
ct
rI
ca
l
co
n
n
ec
tI
o
n
Instructions
Connecting protective ground cable to rail/rails .
In fixed rail systems a protective ground cable must be
installed from the connecting point of the protective ground
cable to the rail .
The following must be observed:
1 . Only connect a protective ground cable to a protective
ground cable connecting point that is approved by an
authorized electrician .
2 . The installation to the mains
must
be done by an
authorized electrican .
3 . The earth cable must be carefully fastened, and a
sufficiently big cable square must be chosen, at least
1 mm² . The installation must be done so that the
resistance from the connecting point of the protective
ground cable to the rail does not exceed 0 .2 Ohm .
Mechanical fixing of the protective ground cable to the rail:
1 . Make a 4 mm threaded hole in the middle of the rail
stripes .
Drill a 10 mm deep hole with a Ø3 .5 mm drill .
Cut a 4 mm thread with a tap .
2 . Place the groundcable with the ringeye .
3 . Mount the ringeye with washer and screw .
4 . The cable must carefully be fixed .
More rails connected to each other:
Normally the assembly brackets will give a
sufficiently low resistance. If this should not
be the case, the rails are connected in series
or in parallel using the same method as
described above. Check with an ohmmeter.
1 Wire, protected ground
2 Washer
3 Screw M4x10 torx
4 Lock washer
Connection to the supply network
Class I
Содержание GH2
Страница 2: ......
Страница 4: ......
Страница 10: ...Bla bla Guldmann GB 1394 11 08 12 ...
Страница 34: ...Bla bla Guldmann GB 1394 11 08 36 ...
Страница 73: ...Bla bla Guldmann GB 1394 11 08 75 4 electrical connection Interconnection diagram GH2 twin ...
Страница 87: ...Bla bla Guldmann GB 1394 11 08 89 ...
Страница 88: ... Guldmann GB 1394 11 08 90 5 control Control 5 Checkliste 88 ...
Страница 90: ... Guldmann GB 1394 11 08 92 6 toolbox ToolS 6 Overview of tools for rail installation 93 ...
Страница 93: ... Guldmann GB 1394 11 08 95 7 service GH2 Service manual ...
Страница 107: ... Guldmann GB 1394 11 08 109 7 service Periodic Inspection ...
Страница 108: ......
Страница 109: ......