Electromagnetic Compatibility
Although the instrument fulfils the relevant EMC requirements precautions
should be taken to avoid unnecessary exposure to electromagnetic fields, e.g.
from mobile phones, etc. If the device is used adjacent to other equipment it must
be observed that no mutual disturbance appears.
Electrical Safety, EMC and Associated Standards
1.
ES 60601-1: Medical Electrical Equipment, Part 1 General Requirements
for Safety
2.
IEC/EN 60601-1: Medical Electrical Equipment, Part 1 General
Requirements for Safety
3.
CAN/CSA-C22.2 No. 60601-1: Medical Electrical Equipment, Part 1
General Requirements for Safety Electrical Equipment for Laboratory
Use
4.
IEC/EN 60601-1-2: Medical Electrical Equipment, Part 1 -
Electromagnetic Compatibility - Requirements and Tests
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions
The GSI Pello is intended for use in the electromagnetic environment specified below. The customer or the user of
the GSI Pello should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - Guidance
RF Emissions
CISPR 11
Group 1
The GSI Pello uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class B Limits
The GSI Pello is suitable for use in all commercial,
industrial, business, hospital, and residential
environments.
Harmonic Emissions
IEC 61000-3-2
Class A Category
Voltage Fluctuations / Flicker
Emissions
IEC 61000-3-3
Complies