![Greyline OCF 6.1 Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/greyline/ocf-6-1/ocf-6-1_user-manual_2264649033.webp)
Page 33
OCF 6.1 Open Channel Flow Monitor
PZxx Series Sensors
Troubleshooting
Resistance measured (between the shield and center wire) across the coaxial cable ends by mulitmeter
indicates ambient temperature.
Resistance vs. Temperature
Values are in Ohms
TEMP
°C
+0
+1
+2
+3
+4
+5
+6
+7
+8
+9
0
10000
10039
10078
10117
10156
10195
10234
10273
10312
10351
10
10390
10429
10468
10507
10546
10584
10623
10662
10701
10740
20
10779
10818
10857
10895
10934
10973
11012
11051
11089
11128
30
11167
11206
11244
11283
11322
11360
11399
11438
11476
11515
40
11554
11592
11631
11669
11708
11747
11785
11824
11862
11901
50
11939
Values are in Ohms
TEMP
°F
+0
+1
+2
+3
+4
+5
+6
+7
+8
+9
30
10000
10021
10043
10065
10086
10108
10130
10151
40
10173
10195
10216
10238
10260
10281
10303
10325
10346
10368
50
10390
10411
10433
10455
10476
10498
10520
10541
10563
10584
60
10606
10628
10649
10671
10692
10714
10736
10757
10779
10800
70
10822
10844
10865
10887
10908
10930
10951
10973
10995
11016
80
11038
11059
11081
11102
11124
11145
11167
11188
11210
11231
90
11253
11274
11296
11317
11339
11360
11382
11403
11425
11446
100
11468
11489
11511
11532
11554
11575
11596
11618
11639
11661
110
11682
11704
11725
11747
11768
11789
11811
11832
11854
11875
120
11896
11918
11939