OCF 6.1 Open Channel Flow Monitor
UNITS/MODE
At
Mode
, press the
and then the
or
to select Flow or Velocity.
Flow mode displays the flow rate in engineering units (e.g. gpm, litres/sec,
etc.) Press the
to store your selection then the
to the next menu item.
At
Linear
press the
key and then the
or
to select your units of
measurement. The Linear units define what units the pipe/channel
dimensions and level reading will be displayed in. Typically inches or mm
is selected. Press the
to store your selection then the
to the next menu
item.
At
Volume
, press the
and then the
or
to select units for volume.
Note: “bbl” denotes US oil barrels. Press the
to store your selection then
the
to the next menu item.
At
Multiplier
, press the
and then the
or
to select the totalizer
multiplier. Multipliers are used when resolution down to single digit is not
required, or when you don’t want to convert from gallons to thousands of
gallons, as an example. Press
to store your selection then
to the next
menu item.
At
Decimal
(Volume), press the
and then the
or
to select the number
of decimal points to be present on the totalizer display on the LCD screen.
Default = 0. Options = 0, 1, 2. Press the
to store your selection then the
to the next menu item.
- - Un i t s / Mo d e - - - - - - - -
Mo d e
Fl o w
L i n e a r
i n
Vo l u me
USG
Mu l t i p l i e r
x 1
De c i ma l
0
s
Fl o w
USG/ m
De c i ma l
2
s
Te mp e r a t u r e
C