Page 59
AVFM 6.1 Area-Velocity Flow Meter
QZ02L Sensor
Velocity Measurement Range:
0.1 to 20 ft/sec (0.03 to 6.2 m/sec) and reverse flow to -5 ft/sec (-1.5
m/sec) in fluids containing bubbles or solids with a minimum size of 100
microns and a minimum concentration of 75 ppm to act as acoustic
reflectors.
Level Measurement Range:
Minimum Head: 1 in (25.4 mm). Maximum Head: 15 ft. (4.57 m)
Operating Temperature:
5 to 175°F (-15 to 80°C)
Exposed Materials:
316 stainless steel, epoxy resin, polyurethane
Sensor Cable:
25 ft. (7.6 m) submersible polyurethane jacket, shielded, 3-coaxial
Sensor Mounting:
includes MB-QZ stainless steel mounting bracket
Temperature Compensation:
Automatic, continuous
1.
62
"
4
1.
2
m
m
0.
56
"
1
4
.2
m
m
MB-QZ - MOUNTING BRACKET
QZ02L VELOCITY/LEVEL SENSOR
25 ft (7.6 m)
Sensor Cable
5.00”
127 mm
1.50"
38.1 mm
0.63"
16 mm
SIDE VIEW
0.
5
0
"
12
.7
m
m
LEVEL
SENSOR
3.00"
76.2 mm
D = 0.2" (5 mm)