![Greyline AVFM 6.1 Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/greyline/avfm-6-1/avfm-6-1_user-manual_2264644030.webp)
Page 30
AVFM 6.1 Area-Velocity Flow Meter
OPTIONAL PZ15-LP LEVEL SENSOR MOUNTING
Mount the PZ15-LP non-contacting ultrasonic level sensor in an unobstructed position at least 8” (203.2
mm) above the high water level. Install the stainless steel mounting bracket in a horizontal position
(check with a level) and then insert the PZ15-LP sensor.
WATER
PZ1 -LP
5
PZ1 -LP MANHOLE MOUNTING
5
PZ1 -LP
5
WATER
PZ1 -LP PIPE MOUNTING
5
QZ02L- VELOCITY SENSOR
B
8",
203.2mm
8",
203.2mm
Содержание AVFM 6.1