ENCLOSURE INSTALLATION
Locate the enclosure within 20 ft (6 m) of the sensor (up to 500 ft -150 m optional). The enclosure can
be wall mounted with the four mounting screws (included) or panel mounted with Option PM Panel
Mount kit from Greyline Instruments.
Avoid mounting the enclosure in direct sunlight to protect the electronics from damage due to
overheating and condensate. In high humidity atmospheres, or where temperatures fall below freezing,
Option TH Enclosure Heater and Thermostat is recommended. Seal conduit entries to prevent moisture
from entering enclosure.
NEMA4X (IP66) WITH CLEAR COVER
1. Open hinged enclosure cover.
2. Insert #12 screws (supplied) through the four enclosure mounting
holes to secure the enclosure to the wall or mounting stand.
Additional conduit holes can be cut in the bottom of the enclosure
with a hole saw or Greenlee-type hole cutter.
DO NOT make conduit/wiring entries into the top of the enclosure.
Note: This non-metallic enclosure does not automatically provide
grounding between conduit connections. Grounding must be
provided as part of the installation. Ground in accordance with the
requirements of the National Electrical Code. System grounding
is provided by connecting grounding wires from all conduit
entries to the steel mounting plate or another point which
provides continuity.
CLEANING
Cleaning of the electronics is not required as a part of normal
maintenance.
The submersible QZ02L sensor may need to be cleaned in
applications with dirty water, or those with a lot of debris.
Use 1 part household bleach to 20 parts water to clean the
sensor. Immerse for 5 hours and then rinse and dry.
Содержание AVFM 6.1
Страница 2: ...Note This page has been left blank intentionally ...
Страница 28: ...Page 28 AVFM 6 1 Area Velocity Flow Meter ...
Страница 34: ...Page 34 AVFM 6 1 Area Velocity Flow Meter Resistance vs Temperature ...
Страница 40: ...Page 40 AVFM 6 1 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ ...
Страница 61: ...Page 61 AVFM 6 1 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration ...