PART FIVE - PRINCIPLE OF OPERATION
See Figure 5-1 for functional diagram pertaining to these
descriptions:
1. The power-supply section (not shown) converts line
power to appropriate voltages for circuit operation.
2. The analog-to-digital converter section accepts signals
from the pH sensor (GLI 5-wire Differential Technique
sensor or combination electrode) and from a tempera-
ture sensor (300 ohm thermistor or 1000 ohm platinum
RTD). This section converts these analog pH and tem-
perature signals to digital signals for use by the
microprocessor.
FIGURE 5-1 Instrument Operations Block Diagram