PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Model 692P
Rev. 9-898
-12-
FIGURE 2-2 Mounting Configurations
2.3 Plugging Conduit
Holes
Use conduit hubs or cable feed-thru fittings where cables
enter the enclosure. Holes not used for cable entry should
be sealed with plugs.
NOTE:
Use NEMA 4 rated fittings and plugs to maintain the
watertight integrity of the NEMA 4 enclosure.
SECTION 3 - ELECTRICAL CONNECTIONS
3.1 GLI 5-Wire Differential
Technique Sensor
To access terminal strips for electrical connections, loosen
bottom four captive fasteners and remove terminal com-
partment cover. Figure 3-2 shows terminal designations for
instrument hookup. If the transmitter is located in a hazard-
ous area, refer to Section 3.5 for wiring details.
NOTE:
For a CE-approved Model 692, the system must be
wired in accordance with GLI hookup drawing
1001X4N1306.
It is recommended that sensor signal wires be run in 1/2"
metal conduit for protection against moisture and mechani-
cal damage. Do not run signal wires in same conduit with
power or control wiring (“electrical noise” may interfere with
sensor signal).
1. Place
SWITCH 3
on back of display module assembly
(Figure 3-2) in
DIFF.
(off/right) position and
TEMP.
COMP
. switch to
THERM.
2. Connect sensor (or interconnect cable) wires to
DIFFERENTIAL SENSOR terminals on TB2 (Figure 3-
2), matching colors as indicated.