PART FOUR - OPERATING AIDS
SECTION 2 - mV CALIBRATION
Rev. 9-898
Model 692P
-47-
SECTION 2 - MV CALIBRATION
2.1 Single-Point Method
2.2 Two-Point Method
If you use the Model 692 to measure the mV output of the
pH sensor for diagnostic purposes, it is recommended to
calibrate the mV measurement using the single or two-point
method for best accuracy.
This method requires a millivolt generator.
1. Disconnect the pH sensor or combination electrode.
2. Connect the millivolt generator in one of the following
ways:
A. When Using GLI 5-Wire Differential Sensor
Connect generator (-) lead to GRN terminal and gen-
erator (+) lead to RED terminal on TB2.
B. When Using Combination Electrode
Connect generator (-) lead to TB4 (REF.) terminal
and generator (+) lead to TB5 (ACTIVE) terminal
post.
3. Set millivolt generator to provide (+) 200 mV.
4. Press
DISP VAR
key as needed to display mV.
5. Press
EXAM/CANCEL
key to place display in “exami-
nation” mode and to indicate mV “CAL VALUE.”
6. The display should indicate “(+) 200 mV”. If not, use
ï
ï
and
ñ
ñ
keys to make display indicate “200 mV.”
7. Press
ENTER
key to enter value (display flashes “OK” to
confirm entry and returns to mV measuring mode).
“HI/LO ERROR” flashes if the difference between the
entry value and actual measured value is greater than
50 mV.
This completes the single-point mV calibration.