PART ONE - INTRODUCTION
SECTION 2 - SPECIFICATIONS
Rev. 9-898
Model 692P
-9-
SECTION 2 - SPECIFICATIONS
2.1 Operational
2.2 Analyzer Performance
(Electrical, Analog
Output)
2.3 Mechanical
Display .........................4-1/2 digit LCD with measurement unit and
setup variable identifiers, 7/8" high digits
Measuring Range:
pH............................0.00 to 14.00 pH
mV...........................(-)500 to (+)500 millivolts
Temperature ............(-)10.0 to (+)110.0
°
C (14 to 230
°
F)
Ambient Conditions ......-30 to 50
°
C (-22 to 122
°
F), 0 to 95% relative
humidity, non-condensing
Temperature
Compensation ..............Automatic, 0-100
°
C (32-212
°
F), 300 ohm ther-
mistor or Pt 1000 RTD, switch selectable
Sensor-to-Transmitter
Distance .......................3000 feet maximum for GLI 5-Wire Differential
Technique sensor; 10 feet maximum for direct
connection of combination electrode (a GLI
Model 714 preamp is required for distances
greater than 10 feet)
Power Requirements ....16 to 40 volts DC
Analog Output ..............Isolated 4-20 mA with output hold feature
Range Expand - The 4-20 mA output can be made to represent a
one pH unit or larger segment of the measuring scale.
Hazardous Area
Classification:
Intrinsic Safety ....... UL and CSA: Class I, Div. 1 Groups A, B, C, & D
Class II, Div. 1 Groups E, F, & G
Baseefa: Zones 0 and 1, Groups IIC through IIA
Division 2............... CSA: Class I, Div. 2 Groups A, B, C, and D
Class II, Div. 2 Groups E, F, and G
♦
Maximum Loop Load
(in series with 692P
and power supply).......With 24 VDC supply: 400 ohms
With 32 VDC supply: 800 ohms
With 40 VDC supply: 1200 ohms
NOTE
- For long cable runs, the resistance of the wire must be
considered and may decrease maximum load capability.
♦
Not applicable when using barrier for intrinsic safety.
Sensitivity.....................0.05% of span
Stability ........................0.05% of span per 24 hours, non-cumulative
Non-linearity .................0.05% of span
Repeatability.................0.01% of span or better
Temperature Drift .........Zero and Span: 0.01% of span per
°
C
Response Time ............1 second to 90% of value upon step change
Enclosure .....................General purpose
safe for Division 2; NEMA
4X, polycarbonate with two 1/2-inch conduit
holes and four stainless steel mounting tabs
Mounting
Configurations ..............Surface mount, optional vertical or horizontal
pipe mounting
Net Weight ...................3 lbs. (1.36 kg) approx.