PART THREE - OPERATION
SECTION 2 - MEASURED VARIABLES
Rev. 9-898
Model 692P
-25-
1.4 Status Indicators
15.
HOLD
indicator (LCD display)
Indicates that the output hold feature is in use (Model
692 output value is maintained).
NOTE:
After 30 minutes,
HOLD
indicator flashes to in-
dicate that output hold feature will be
automatically cancelled in 30 seconds. Pressing
OUTPUT
key extends hold feature for another
30 minutes.
16.
OK
indicator (LCD display)
Flashes for approximately 5 seconds to confirm suc-
cessful entry of a setup variable value.
17.
LOCK
indicator (LCD display)
Indicates that Model 692 keypad entry is “locked” to
prevent unauthorized alteration of stored setup variable
values. Refer to Section 8 for security lock feature in-
structions.
NOTE:
When the 692 is locked, you can still enter cali-
bration values and display stored entry values.
18.
ERROR
indicator (LCD display)
Flashes to indicate an incorrect entry or alternately
flashes with “Err 1”, “Err 2”, “Err 3” or “Err 4” to indicate
a system diagnostic error causing improper system op-
eration.
SECTION 2 - MEASURED VARIABLES
The Model 692 can display four measured variables. With
the display in the measurement mode, each press of the
DISP VAR
key sequentially displays:
n
pH.
n
Temperature in
°
C or
°
F.
n
The sensor's mV signal.
n
The 4-20 mA instrument analog output.