PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Rev. 9-898
Model 692P
-15-
Indirect Hookup
with Model 714 Preamp
(distances more
than 10 feet)
3.3 Power Supply
3.4 Analog Output
1. Place
SWITCH 3
on back of display module assembly
(Figure 3-2) in
DIFF.
(off/right) position and
TEMP.
COMP.
switch to
THERM.
2. Locate the GLI Model 714 preamp within 10 feet of the
combination electrode. Refer to the Model 714 preamp
instruction manual for electrical connection details be-
tween the combination electrode and the preamp.
3. The Model 692 can be located up to 3000 feet from the
Model 714 preamp. Connect interconnect cable from the
preamp to the Model 692 DIFFERENTIAL SENSOR ter-
minals on TB2 (Figure 3-2), matching colors as
indicated.
Connect the DC voltage power supply to “4-20 mA” termi-
nals on TB1, matching polarity as indicated.
NOTE:
If the 692 is used in an intrinsically safe application,
it may be located in a Class I or II, Division 1 haz-
ardous area without an explosionproof enclosure
when powered through an approved barrier. Refer
to Section 3.5 for details on hazardous area wiring
requirements.
The isolated 4-20 mA output can represent either the
measuring scale or a selected segment of it. To use the
range expand feature, refer to Part Three, Section 7.1 for
details.
The isolated 4-20 mA output can drive auxiliary devices (re-
sistive loads) such as displays, recorders and computers,
provided that the voltage supplied by the power supply is
adequate. Devices must be wired in series with the trans-
mitter and power supply. The voltage drop across the
load(s) and the 16 volts DC minimum needed to drive the
transmitter determines the minimum voltage required from
the power supply.
1. Determine the necessary voltage required to adequately
drive the Model 692 and auxiliary device(s).
A. The Model 692 acts as a current controlling device.
Thus, the current output remains the same even if
the power supply voltage fluctuates or the load re-
sistance changes. The current varies only with
respect to the sensor signal, as long as the voltage
Содержание 692P