PART THREE - OPERATION
SECTION 6 - pH CALIBRATION
Rev. 9-898
Model 692P
-39-
Selecting
This Method
Entering Calibration Points
(by qualified person)
1. With display in pH measuring mode, press
EXAM/
CANCEL
key to place display in “examination” mode.
2. Press
NEXT
key until display indicates “BUFFER CAL.”
3. Press
ï
ï
key to make display indicate “tAbL.”
4. Press
ENTER
key to select “two-key” table method
(“OK” flashes to confirm entry).
# # #
1. The following table lists the selection of pH values and
their buffer formulations. Choose and note one or two
table pH values for use as the calibration point(s).
NOTE:
When two calibration point values are selected,
they must be at least one table value apart from
each other (non-adjacent values).
Table E
BUFFER VALUES FOR
“TWO-KEY” TABLE METHOD
pH Buffer Value
Buffer Formulation
1.09
HCl (DIN 19267)
1.68
Tetraoxalate (NBS 260-53)
3.56
Tartrate (NBS 260-53)
3.78
Citric acid (DIN 19266)
4.00*
Pthalate (DIN 19266)
4.01
Pthalate (NBS 260-53)
4.65
Acetate (DIN 19267)
6.87
Phosphate (DIN 19266)
7.00
Phosphate
7.41
Phosphate (DIN 19266)
9.18
Borax (NBS 260-53)
10.00 (10 A on display)*
Carbonate/bicarbonate
(0.5 Molar)
10.00 (10 b on display)
Glycol
12.45
Calcium hydroxide (DIN 19266)
*Indicates default values for LO and HI BUFFER VALUES.
2. Press
NEXT
key until display indicates “LO BUFFER
VALUE.”
3. Use
ï
ï
and
ñ
ñ
keys to make display indicate the de-
sired value for the single or lower calibration point.