PART TWO - INSTALLATION
SECTION 1 - UNPACKING
Model 692P
Rev. 9-898
-10-
PART TWO - INSTALLATION
SECTION 1 - UNPACKING
After unpacking, it is recommended to save the shipping
carton and packing materials in case the instrument is
stored or re-shipped. Inspect the equipment and carton for
damage. If there is evidence of shipping damage, notify the
carrier immediately.
SECTION 2 - MECHANICAL REQUIREMENTS
2.1 Location
2.2 Mounting
An agency approved Model 692 is designed intrinsically
safe. That is, an explosionproof enclosure is not required
when the transmitter, powered through an approved barrier,
is located in Division 1 hazardous areas.
WARNING: THE POWER SUPPLY AND INTRINSIC
SAFETY BARRIER MUST ALWAYS BE LOCATED IN A
SAFE AREA.
1. Locate the 692 within 3000 feet of where the GLI 5-wire
Differential Technique sensor is to be installed. If a
combination electrode is used, the 692 must be within
10 feet of the electrode for a direct cable run. A GLI
Model 714 preamp may be used to extend this distance
to 3000 feet, but the preamp must be within 10 feet of
the electrode.
2. Mount the 692 in a location that is:
n
Clean and dry where little or no vibration exists.
n
Protected from falling corrosive fluids.
n
Within ambient temperature limits (-22 to 122
°
F,
-30 to 50
°
C).
CAUTION: MOUNTING IN DIRECT SUNLIGHT MAY
INCREASE TEMPERATURE ABOVE MAX. LIMIT.
Refer to Figure 2-1 for enclosure and mounting dimension
details. Figure 2-2 illustrates various mounting configura-
tions. Use the four stainless steel tabs to surface-mount the
instrument. An optional hardware kit is required for pipe-