GL820-UM-251-9370
4-13
4. SETUP PROCEDURES
Setup procedure
(1) As shown in the figure above, connect the reference resistor to the GL820’s input terminals.
(2) Leave the GL820 turned on for at least
30 minutes
. And leave the GL820 turned on for at least
10
minutes
after wiring.
(3) After entering the System Setup mode, open the Setup menu as shown below.
Specify the AMP Setup conditions as shown below.
Mode
:Setup3 (RTD temp.)
CH
:CH 20
(4) Move the cursor to the “Execute Setup” position as shown below.