Device Description HG G-19370-B/HG G-19380-B | English, Revision 03 | Date: 05.05.2020
23
Configuration – Chapter 6
6.4.5 (4) CSV
Serial output of the sum and difference signals as Comma Separated Values (CSV).
These values can be written to a file via the terminal program. Those files can then
be imported into a spreadsheet application like Microsoft® Excel® and then evalu
-
ated. As long as the sum voltage is below the threshold 0 is output. The values for
the induced voltages are output in mV as shown in Figure 2 on page 9.
Figure 15
Example for a CSV output
In case you want to have values output at all times you can temporarily set the De
-
tect threshold in the menu Calibration Config to 0.
Table 10
Meaning of the values output in the menu Serial Data-Stream
Freq. 1
Freq. 2
Freq. 3
Freq. 4
Freq. 5
Energy Track Encoder
Sum1 Diff1 Sum2 Diff2 Sum3 Diff3 Sum4 Diff4 Sum5 Diff5 Sum6 Diff6 Incr. Enc.
CAUTION
Invalid Detect Signal
A threshold of 0 means that the sensor permanently detects a track. Then
Detect
Signal
can no longer be used as an aid to detect whether the vehicle leaves the
track.
Only ever set this threshold to 0 for testing purposes.