B l u e B o x – M e n u O p e r a t i o n
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 77 / 96
www.go-sys.de
8.2.7 Current Output
Sensor list 8.2.2
The BlueBox has two galvanically isolated current outputs, see also
1.5
Notes on the Current Outputs
.
Current output 1
bsa 503-A | Current output 2
bsa 503-B
A current output is an analogue actuator and is parameterized similar
to a sensor. The measurement values of an assigned sensor control a
current output from 4 mA to 20 mA.
The assignment of a sensor and the parameterization of the
current output are usually made with the program AMS
1
.
Example mg/l to mA:
Menu I
Switches to the input of calibration values.
2
Switches to the inpu of a calibration mode.
2
see
8.2.2.1.1 Calibration
Selects the current output for the multiple parameter display or not.
see
8.2.1.6 Display
Switches to the table display of the current values. see 8.2.2.1.2
Switches to the diagram display of the current values. see 8.2.2.1.3
Switches to the Info menu of the current output.
Switches back to Sensor list.
1
The program AMS is part of the BlueBox PC Software.
2
Except in special cases, calibration is performed with the AMS program.
bsa 503-A
Minimum value
Measuremet range
Maximum value
Calibration
Calibration mode
selected
selected
Table
Diagram
Overview
<–
bsa 845-A
<–
selected
Calibration
Calibration mode
Table
Diagram
Overview
bsa 845-A
<–
Name
Unit
Parameter
Interval
Average
Resolution
bsa 845-A
<–
Lower limit
Upper limit
Menu III
Menu II
Menu I