Appendix J – Status Messages
Status no. Description
Display
0
Sensor sends data.
Measurement value
1
Sensor sends no data.
No Data
2
A new sensor is recognized. (temporary at sensor initialization)
Wait
3
Sensor-ID assigning. (temporary at sensor initialization)
Wait
4
Measurement value is unreliable.
(currently only for spectrometers)
[
Measurement value
]
30
Formula error
Error at line
n
31
Unknown sensor is used in the formula.
? Sensor
33
Default calculation time for for and while loops is exceeded.
Calc Timeout
50
Minimal measurement value underrun (virtual sensor)
<
minimal measurement
value
51
Maximal measurement value overrun (virtual sensor)
>
maximal measurement
value
52
Internal communication error
COM
53
Underrun of the lower limit of the AD converter
ADC min.
54
Overrun of the upper limit of the AD converter
ADC max.
55
General device error
Dev Error
The entries in this column can be queried using AMS Formula.
The entries in this column appear as status messages in the parameter display.
In 6-way parameter display (see
8.1.2 Parameter Display 6-way
) status messages are displayed shortened.
In 8-way parameter display (see
8.1.3 Parameter display 8-way (Tiles)
) no status messages are displayed, but
highlighted in colour: yellow
Status no. 0 | orange
Status no. 4, 50, 51 |
red
all others |
There is also a status message with the status number 57, which refers to an exceeded calibration interval of
the clear water calibration of a spectrometer and is marked orange in the tile display.