MaxiMet
Page 27
Issue 1
Doc. No. 1957-PS-000x
March 2015
6.
CONFIGURING WITH METSET
MaxiMet can be configured using Gill Instruments MetSet Software which is loaded on the CD
supplied with MaxiMet.
MetSet software can run on PC’s running up to and including Windows 8 and can also be
downloaded from the Gill Website:-
http://www.gillinstruments.com/main/software.html
.
6.1.
MaxiMet Default Configurations
GMX200 Factory Default Data String:
<STX>Q,127,000.03,089,
2000-01-01T00:40:50.2,+10.5,0000,<ETX>40 & (CR,LF)
STX
Node letter
Wind direction
Wind speed
Corrected Wind Direction
Date and Time
Supply voltage
Status code
ETX
Checksum
Carriage Return/Line Feed
NOTES:
<STX> is the Start of String character (ASCII value 2).
<ETX> is the End of String character (ASCII value 3).
Checksum, the 2 digit Hex Checksum sum figure is calculated from the Exclusive OR of the
bytes between (and not including) the STX and ETX characters.
Q,127,000.03,000,2000-01-01T00:40:50.2,+10.5,0000, 21
NOTE.
Switch on Time
.
Upon switching on the MaxiMet, allow 9.5 seconds to elapse before data is output.