MaxiMet
Page 56
Issue 1
Doc. No. 1957-PS-000x
March 2015
Safe Mode Method 2
Connection
Connect the MaxiMet for RS232 communication as detailed in Para 4.6.
Note at this stage the MaxiMet supply to be switched off.
Open a Terminal program e.g. Gill WIND Software (obtainable from
http://www.gillinstruments.com/main/software.html
) or equivalent terminal program e.g.
Tera-Term, Putty etc.
If using Gill Wind Software note that the Tools features are not applicable.
Open Gill Wind Software.
Serial Port
:
Set the drop down menu to the required COM Port Connection.
Click on the
OK
button.
Baudrate
:
Set the drop down menu to
4800
Bauds.
A blank Wind Terminal screen will be opened.
Left Click the mouse in the blank terminal screen.
Hold down the * key on the keypad.
With the * key still held down, apply power to the MaxiMet.
This will result in placing the unit into SAFE MODE.
Press Enter to start a new line (Ignore illegal command line).
To now change communication from say SDI-12 (or MODBUS) or RS422 or RS485 to Default
RS232.
Type COMMS RS232 and press Enter.
Type Q and press Enter to exit SAFE MODE.
Data will now scroll on screen.