Mounting and operating manual GIA 20 EB with electrically insulated supply
page 30 of 32
Err.4:
Values below display range
Indicates that display value is below the valid display range of the device (-1999 digit).
Possible causes:
- Incorrect scale.
- Counter underflow.
Remedies:
- The error-message will be reset if the display value is above -1999.
- Reset the counter
– When happening frequently, check the scale-setting, maybe it was set too
low and should be increased.
Err.7:
System-error
The device features an integrated self-diagnostic-function which checks essential parts of the de-
vice permanently. When detecting a failure, error-message Err.7 will be displayed.
Possible causes:
- Valid operating temperature range has been exceeded or is below the
valid temperature range.
- Device defective.
Remedies:
- Stay within valid temperature range.
- Exchange the defective device.
Err.9:
Sensor defective
The device features an integrated diagnostic-function for the connected sensor or transmitter.
When detecting a failure, error-message Err.9 will be displayed.
Possible causes:
- Sensor broken or sensor shorted (Pt100 or Pt1000).
- Sensor broken (thermo-elements).
Remedies:
- Check sensor or exchange defective sensor.
Er.11: Value could not be calculated
Indicates a measuring value, needed for calculation of the display value, is faulty or out of range.
Possible causes:
- Incorrect scale.
Remedies:
- Check settings and input signal
13. Decommissioning, reshipment and disposal
13.1. Decommissioning
Always disconnect the device before from its supply before decommission (e.g. at fuse). Valid
general safety requirements shall be observed.
Please also make sure that connected Loads are disconnected also and are in a safe state.
13.2. Reshipment and disposal
All devices returned to the manufacturer have to be free of any residual of measuring
media and other hazardous substances.
Measuring residuals at housing or sensor may be a risk for persons or environment
Use an adequate transport package for reshipment, especially for fully functional
devices. Please make sure that the device is protected in the package by enough
packing materials.
Add the completed reshipment form of the GHM website
http://www.ghm-messtechnik.de/downloads/ghm-formulare.html
The device must not be disposed in the unsorted municipal waste! Send the device
directly to us (sufficiently stamped), considering the above if it should be disposed.
We will dispose the device appropriate and environmentally sound.