Instructions for Use
CPET
Version: 4 | Release date: 01 April 2022
page 97
Error
Rectification
Replace filter and Permapur
Carry out gas calibration again
Check the device for faults, see the
IFU of the device
Other obvious measurement error
Check the fit of the mask / noseclip
Make sure that the patient tightly
encloses the mouthpiece with his or
her lips during the examination
Check connection between device
and flow sensor
Ensure correct execution of the
breathing manoeuvre
Check entered environmental
conditions
Perform volume calibration again
Carry out gas calibration again
Check the device for faults, see the
IFU of the device.