Error
Rectification
No heart rate
Check correct fit of electrodes / chest
strap / SpO
2
sensor
Check ECG signal for quality
Check interface setting to ECG
software
Error message "Significant deviation
during zero adjustment detected".
Carry out a new gas calibration
Ventilate the examination room
Ensure air circulation around the
device
Check the device for faults, see the
IFU of the device
Error message "The determined
delay does not match the entered
tube length".
Correct the set tube length
Check tubes for moisture
Replace filter and Permapur
Check the device for faults, see the
IFU of the device
Measurement error: VO
2
, VCO
2
and
VE too low
Check the fit of the mask / noseclip
Make sure that the patient tightly
encloses the mouthpiece with his or
her lips during the examination.
Check connection between device
and flow sensor
Check tubes for moisture
Check entered environmental
conditions
Perform volume calibration again
Check the device for faults, see the
IFU of the device
Measurement error: VO
2
and VCO
2
too low, VE normal
Check connection between device
and flow sensor
Determine zero point again