Instructions for Use
CPET
Version: 4 | Release date: 01 April 2022
page 95
Error
Rectification
Button for starting the measurement
is missing
Check the connection of the device to
the PC
Check the rights of the logged-in user
When calling via an interface (GDT,
HL7) check the requested
examination type
Measurement ends automatically
after a short time
Check interface setting to ECG
software
Examination does not start,
graphics remain stored
Check interface setting to ECG
software
No flow signal
Check correct connection of the
device
Check connection between device
and flow sensor
Check tubes for moisture
Check the device for faults, see the
IFU of the device
Drift on volume signal
Determine the zero point of the device
again
Check entered environmental
conditions
Check connection between device
and flow sensor
Check the device for faults, see the
IFU of the device
No O
2
signal or CO
2
signal
Check connection between device
and flow sensor
Check tubes for moisture
Check the device for faults, see the
IFU of the device