Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Version: 7 | Release date: 02. August 2021
Page 81
Reason: Electric shock. Triggering malfunctions of
Bodystik. Unqualified user / operator can detect
sources of failures too late or even cause them.
Therefore:
•
All maintenance and servicing work may only be
performed by specialised personnel who have
been authorised by the manufacturer!
8.2.1
Checking / Exchange of Tubes and Cables
All parts of Bodystik should be checked for visible mechanical
damage (cracks, tears) each day. If damage is ascertained, the
corresponding component must be replaced.
For replacing components see:
•
Chap. 4.1.3 “Connections / Interfaces of the Bodystik”
•
Chap. 4.1.4 “System Construction and Electrical
Safety”
•
Chap. 6.2.2 “Recommissioning after Maintenance /
Cleaning Work by the Operator”