Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Page 80
Version: 7 | Release date: 02. August 2021
Attention
Independent of this, the user / operator must carry out regular
checks during day-to-day operation, see also the following
chap. 8.2 “Servicing / Maintenance by the User / Operator”!
8.2
Servicing / Maintenance by the User / Operator
In order to ensure a flawless operation of Bodystik over its whole
service life, regular servicing and repairs, if applicable, are
required.
Interval
Servicing work
Once per day
Visual checks of the device and its components for
damage and replacing them if necessary
(see chap. 8.2.1 “Checking / Exchange of Tubes and
Cables”)
Once per day
Checking the O-Rings and Sealing Rubbers for tears /
wear and replace if necessary
(see chap. 8.2.2 “Checking / Replacing O-Rings and
Sealing Rubbers”)
Once per week
Brush the door rubber with Vaseline
(see chap. 8.2.3 “Care of Door Rubbers”)
Once per year
Replacing the O-Rings and Sealing Rubbers
(see chap. 8.2.2 “Checking / Replacing O-Rings and
Sealing Rubbers“)
Possible danger to life
Reason: electric shock. Misdiagnosis caused by
damaged components of the Bodystik.
Therefore
•
Always make controls before starting the
Bodystik examination!