Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Page 34
Version: 7 | Release date: 02. August 2021
3.2
The Technical State of Bodystik and System Construction
Possible danger to life.
Reason: Electrical shock. Cross-contamination or
misdiagnosis. For this: Only use Bodystik within its
performance limits! Therefore:
•
Do not overstress Bodystik Use with care!
•
Do not modify or use product Bodystik or
complete system contrary to the respective
manufacturer's specification!
•
Only use Bodystik within its expected service life,
determined by the manufacturer!
•
Under no circumstances use or connect any
devices, systems, equipment and other products
that are not part of the complete system!
•
Never obstruct the access to the mains plug or
On / Off switch of the equipment cart!
Disconnection of the power supply must be easily
accessible!
•
Only use accessories and consumables which are
authorised by the manufacturer as well as
authorised replacement components, add-ons
and auxiliary materials.
Check that the components are in a functional
and safe condition!
•
Ensure that in the patient environment – a
distance of 1.5 meters to the patient – there are
no accessible electrical parts (interfaces, plugs,
etc.) that are not isolated from the mains with an
isolation voltage of 4 kV!