Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Version: 7 | Release date: 02. August 2021
Page 37
Reason: Disregard of a contraindication. Misdiagnosis
caused by measurement error. Cross-contamination.
For this: Observe the general medical principles!
Therefore:
•
Inform yourself and observe the respective
contraindications before each test!
•
When carrying out tests, observe the content of
the applicable guidelines and recommendations
(e.g. ATS / ERS Guidelines)!
Reason: Collapse of the patient due to claustrophobia.
For this: Observe general medical principles! Therefore:
•
Before examination, inform the patient about the
emergency opening of the door!
•
Do not leave the patient unattended in the cabin!
Possible severe physical injury.
Reason: Bodystik can tilt over. Therefore:
•
Observe the maximum weight of the person on
the chair when chair is swivelled outwards. It is
125 kg!
Possible physical injury
Reason: Skin burn caused by short-circuit inside the
device. Therefore:
•
Do not touch shutter drive in case of excessive
heat!