Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Version: 7 | Release date: 02. August 2021
Page 89
9.2.3
Shutter System
9.2.3.1
Shutter Cage or Shutter Block
1.
Disassemble the shutter system
For this:
Procced in reverse order as described
in chap. 6.2.3.1 ”Assembly Shutter
System“.
2.
a) Until product. 11 / 2018:
disconnect the shutter block [1] from
the shutter drive [2]
by turning it counterclockwise.
Also separate the shutter block [1]
from the shutter plug [3] by screwing
and pulling it out.
b) Until product. 11 / 2018:
disconnect the shutter cage [4] from
the shutter drive [2]
by turning it counterclockwise.
3.
Remove visible dirt
with a soft cloth.
4.
Put the components [1] + [3] or [4] +
[5] in disinfectant fluid.
5.
Wash the components [1] + [3] or [4]
+ [5] throughly in distilled water
so that no residues of contamination
and disinfectants remain.
6.
Components [1] + [3] or [4] + [5]
air dry
until no more moisture is visible.
4
5
4
1
3
2
2