Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Page 98
Version: 7 | Release date: 02. August 2021
12.2
Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Bodystik
guaranteed by the manufacturer and for the safety of the patient
and user / operator.
Storage / Transport:
min.
max.
Temperature:
-10 °C
+50 °C
Relative air humidity:
0 %
95 %
non-condesing
Operation:
min.
max.
Environmental temperature:
+10 °C
+35 °C
Avoid extreme
temperature fluctuations!
Relative air humidity:
20 %
95 %
non-condensing
(at least 30 % for synthetic flooring)
Atmospheric pressure:
800 hPa
1080 hPa
Setup:
fixed
Space requirement cabin:
(door completely open)
(L) 1600 mm x (W) 950 mm x (H) 1730 mm
Necessary ceiling load:
(cabin + patient)
300 kg/m²
Flooring:
Wood, concrete or ceramic tiles
stable; even; conductive
Environment / room:
Closed, clinical area
air-conditioned
not an explosive or flammable environment!
not near an MRI!
not in direct sunlight
not draught / flow to air conditioning system
well ventilated
Operating Mode:
Continuous operation