Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Page 60
Version: 7 | Release date: 02. August 2021
6.2.1
Recommissioning after Termination by Patient
If the door is opened by the patient from the inside of the cabin,
the Bodystik still remains functional.
Only the current measurement is aborted.
You can now decide if the interrupted measurement should be
resumed or continued with the next measurement.
Follow the Software instructions.
To continue the "Resistance" and "TGV" measurements, the
door must be closed again. If this is not done, the
BLUE CHERRY® Software will inform you with the message
"Please close the door"
You can find more information on:
•
Resistance and TGV
in chap. 4.2.2 "Door-Opener Patient Panel“
•
Software operation generally
in the separate IFU BLUE CHERRY®
6.2.2
Recommissioning after Maintenance / Cleaning Work by the Operator
Before the Bodystik can be used to perform measurements
again, all components must be properly reconnected, the USB
connection to the computer must be established, and the
Bodystik must be connected to the power supply and calibrated.
Calibration also fulfils the regulatory requirements for functional
testing after maintenance work.
Please proceed as described in the respective chapters.