PSY-201 User Guide
GP-UM-PSY-201-11
Page 90 of 105
Creation of a new calibration matrix requires measurement of at least 50 different SOPs.
Click “OK” on the notification window button.
Click “Start” to begin matrix generation. PolaView uses the PSY-201’s triangle scrambling
function at a scrambling rate of 3
×
2
π
/s to create the required set of diverse SOPs.
The points will be marked on the sphere as they are recorded, and the status window will
notify the user when the requisite number of points has been recorded.