:OUTPut:ANALog<n>:MIN <float> Set
output parameter min value for channel n.
n = 1 to 4.
Range: DOP: 0 to 100%
POW:
−
60 to +15 dBm
S1, S2, or S3:
−
1.00 to +1.00
dREF: 0 to 180.0°
MIN < MAX
:OUTPut:ANALog<n>:VLEVel <float>
Set
analog output voltage max level.
Range: 0 to 5V.
Default: 5V.
Polarization Control
:CONTrol:SOP <float1>,<float2>,
<float3>
Set SOP by Stokes parameters and enable tracking.
Stokes parameters are auto-normalized to unit vector
before tracking:
SOP will be set to (f1/A, f2/A, f3/A) where
A=sqrt(f1*f1, f2*f2, f3*f3)
Note: f1=f2=f3=0 is not allowed.
Example:
:CONT:SOP f1,f2,f3
:CONTrol:ANGLe
<float1>,<float2>
Set SOP in spherical coordinates and enable tracking.
Ranges:
f1=theta: 0 to 360
f2=phi: 0 to 180
Unit: Degrees.
:CONTrol:ANGLe:ELLipse
<float1>,<float2>
Set SOP in ellipse angles and enable tracking.
Ranges:
f1=azimuth angle:
−
90 to +90 f2=ellipticity angle:
−
45 to +45.
Unit: Degrees
Relation to Stokes parameters:
S1=cos(2*ELPA)*cos(2*AZM);
S2=cos(2*ELPA)*sin(2*AZM), S3=sin(2*ELPA)
:CONTrol:DISable
Stop active tracking.
:CONTrol:ENABle
Track to current SOP.
Equivalent function to TRACK front panel quick key.
:CONTrol:TSCan:TYPE <integer>
Set
axis
of rotation for trace scan function:
Type:
0: LP0
1: LP45
2: LP90
3: LP135
4: RHC
5: LHC
Default: 0
:CONT:TSC:TYPE 1
:CONTrol:TSCan:STEP
<float>
Set trace scan step size
Unit: degrees, Range: 0 to 180
Default: 1
:CONTrol:TSCan:STEP 90