PSY-201 User Guide
GP-UM-PSY-201-11
Page 36 of 105
Default Angle Unit: degrees
Amplitude
The amplitude unit is the unit used to display the amplitude of the signals applied to the
retarders of the polarization control unit inside the PSY-201. The amplitude can be displayed in
volts, radians, or multiples of
π
radians. Radians or
π
units are calibrated to an accuracy of ±3%.
The amplitude unit setting applies to the waveform amplitude and offset in modulation
mode and to the settings for each fiber squeezer in manual control mode.
Default Amplitude Unit: Volts
Integration Time
Integration time refers to the detector integration time used for each SOP sample. The
integration time can be set for either automatic or manual control. In automatic control, the
integration time is dynamic, with its value determined by the gain, which can change with input
power level.
Gain
3dB bandwidth
Average samples
Integration
time
0-5 1MHz
4
1µs
6,7,8 600kHz
8
2µs
9,10,11 100kHz
40
10µs
12-18 10kHz
400
100µs
Default setting: Auto
Amplitude Unit
Voltage
SET
L
Integration Time
Auto
SET
L
Angle Unit
Degree
SET
L