PSY-201 User Guide
GP-UM-PSY-201-11
Page 46 of 105
Ethernet
1.
Make sure that the Ethernet communication setting (Static or Dynamic) matches the
Ethernet addressing type being used.
2.
Make sure that the cable type is correct: cross-linked cable for static IP with direct
connection between computer and instrument, straight-linked cable for connection via a
router.
3.
Every time “Dynamic IP” is selected from the Settings menu, the instrument requests a
new dynamic IP address from the server. If this happens while the instrument is already in
communication via dynamic IP, the communication may be interrupted.
PolaView
1.
Make sure that the PSY-201 “Command Type” is set to SCPI and the EOS is set to
<CR+LF>. PolaView will not work with other command settings.
2.
Connect the PSY-201 to the control computer via USB. PolaView does not work with other
communication interfaces.
3.
Stop measurement before changing measurement setup or parameters. In many cases,
PolaView will not respond to menu selections if a measurement is running.
4.
Stop measurement and/or close PolaView before powering off the PSY-201.