PSY-201 User Guide
GP-UM-PSY-201-11
Page 62 of 105
*TRA:SCN?
Query the trace scanning speed
(dwell time per point, in ms)
*TRA:SCN 10# (dwell time = 10 ms)
*TRA:STE n.f#
Set the trace step size, in degrees
Default 1°
Range: 0 to 180
*E00# if successful else *Enn# error
message
*TRA:STE?
Query the step size
*1.5#
*TRA:STA?
Query the trace scan axis
N= 0: 0°
N= 1: 45
N= 2: 90
N= 3: 135
N= 4: RHC
N= 5: LHC
Range 0 to 5
*TRA:STA 0#
*TRA:STA N#
Set the trace scan axis
N= 0: 0°
N= 1: 45
N= 2: 90
N= 3: 135
N= 4: RHC
N= 5: LHC
Range 0 to 5
*E00# if successful else *Enn# error
message
*RAN:ENA? Query
the
discrete scrambling
enable/disable status.
*RAN:ENA 0|1#
0 = disabled, 1 = enabled
*RAN:FRQ? Query
the
discrete scrambling
frequency.
*RAN:FRQ 10.00 #
*RAN:FRQ nnnn
#
Set the discrete scrambling frequency.
Range: 0.01 to 20000 points/s
*E00# if successful else *Enn# error
message
*RAN:ENA ON|OFF#
*RAN:ENA 1|0#
Start or stop discrete scrambling.
1 or ON = enabled.
*E00# if successful else *Enn# error
message
*SAW:FRQ?
Query the triangle scrambling
frequency.
Range: 0.01 to 2000
Units: multiples of 2
π
/s
*SAW:FRQ 300.10#
*SAW:FRQ FFF.F#
Set the triangle scrambling frequency
Range: 0.01 to 2000
Units: multiples of 2
π
/s
*E00# if successful else *Enn# error
message
*SAW:ENA?
Query the triangle scrambling
enable/disable status.
*SAW:ENA 0|1#
0 = disabled, 1 = enabled
*SAW:ENA ON|OFF#
*SAW:ENA 1|0#
Start or stop triangle scrambling.
1 or ON = enabled.
*E00# if successful else *Enn# error
message
*MCV n xxx.x#
Set manual control voltage on channel
n
n: 1 to 4
xxx.x: 0-150.0
xxx.x unit: V
*E00# if successful
else *Enn# error message
Communication
*COM:TRG ON|OFF#
Enable/disable external measurement
trigger
*E00# if successful else *Enn# error
message
*GAN
?
Query gain value
*6#
*RSX?
Query RS-232 Baud Rate
*9600#
Note on operation modes: ANG mode includes both spherical coordinates and ellipse angles.