The REF button allows the user to set a reference state (SOP, DOP, optical power), which
can be used to monitor changes in these parameters. When the REF button is pressed while there
is SOP information on the OLED display, e.g. from the measurement screen or the SOP Storage
screen, the instrument will ask whether to set the displayed information as the reference.
Select “Yes” to proceed.
If “No” is selected, or if the REF button is pressed while there is no SOP information
displayed on-screen, then the reference menu is shown:
1.
“Show Current Ref.” displays the current SOP, DOP, and Power reference values.
Press the ENTER or REF buttons to return to the reference menu.
2.
“Set to Basis Ref.” allows the user to set one of the basis polarization states as the
reference. Use the arrow keys
▼▲►◄
to select the desired state, and press the
ENTER button to confirm the selection. Press the REF button to return to the menu
without making a selection.