PSY-201 User Guide
GP-UM-PSY-201-11
Page 39 of 105
Figure 21 dREF trigger illustration
With the dREF output trigger enabled, the PSY-201 outputs TTL low as long as the
measured SOP is within the blue circle, and TTL high when the SOP is outside of the blue circle.
The dREF output trigger signal uses pin 15 on the DB15 expansion port on the rear panel
of the PSY-201 (Figure 7).
Use
▲▼
to change the value of the underlined parameter or digit, then press ENTER to
confirm the selection and move the cursor to the next parameter. An “OK” indicator on the right
side of the screen shows that the field has been updated.
◄►
can be used to move the cursor
without updating the field.
Enable/disable
status:
Default:
disabled
Threshold Range: 0 to 180° from reference SOP on Poincaré sphere
Default: 0°
TTL level low = within threshold
TTL level high = beyond threshold
Communication
This submenu allows the user to configure remote control options. In general, we
recommend using the front panel, rather than remote control commands, to configure
communication options.
Circle defined by dREF threshold
Reference SOP
Trigger Out: Enable
Threshold: 000.0°
SET
OK
L