Model IR5500
56
8.2 Alignment Using Legacy AJ Value
To change the alignment signal strength indication from Pct to AJ, apply the magnet over the
GM logo on the Receiver cover until rSt appears. When SE appears, remove the magnet.
Apply the magnet when diS appears. When Pct appears, apply the magnet until AJ appears.
Remove the magnet. Apply the magnet when Fi appears. The display will cycle through more
menu choices. When Fi appears again, apply the magnet to exit the setup menu. For a
diagram of the menu choices, refer to the User Menu in the Operation section.
During alignment, verify that the AJ level is comparable to typical values listed below. For
80150-meter configuration, add 50 meters to the range.
Range
20 m
30 m
40 m
50 m
60 m
70 m
80 m
90 m
100 m
AJ value
72 to 67
65 to 60
59 to 54
53 to 48
49 to 43
45 to 38
43 to 35
39 to 32
35 to 30
Typical AJ Signal Level Values