Model IR5500
43
5.1.3 F3 Beam Block
This fault indicates that IR input is below the Low IR fault threshold.
ACTION
– The system should be aligned and the optical path should be
checked for blockage or windows cleaned.
5.1.4 F4 IR Flash Timing
Flashes from the Source are erratic.
ACTION –
Verify that IR is only from one Source. .
5.1.5 F5 Setup Menu
This fault indicates that the user has not exited the setup menu after 6 minutes
of inactivity.
ACTION –
Apply the magnet to exit the setup menu.
5.1.6 F6 Low Voltage Input at the Receiver
ACTION
– Ensure the supply voltage at the Receiver is within specification..
5.1.7 F7 Code Checksum
This fault indicates a fault in device program memory.
ACTION
– Cycle power. If the fault persists, contact the factory.
5.1.8 F8 Fault during Zeroing
This fault indicates that a fault occurred during zeroing.
ACTION
– Repeat zeroing with no faults present.
5.1.9 F9 Gas Left
This fault indicates the unit timed out while in Gas Check Mode.
ACTION
– Remove the Test Gas Film and apply the magnet to clear the fault.
5.1.10 F10 Reset Short
This fault indicates that the reset external connection is shorted to ground for
more than 30 seconds.
ACTION
– Remove the reset short wire from ground.