Chapter 7 - Diagnostics/Troubleshooting
7-23
Section 7-8
Error Messages
NOTICE
If the problem (error message) still exists after performing the described actions, call technical support.
Error Messages
Actions
3D B_RxLines exceeded!
restart the system
3D B-O_Frames exceeded!
restart the system
3D C_RxLines exceeded!
restart the system
3D C-O_Frames exceeded!
restart the system
530-Probe connected on……….
Disconnect and reconnect the probe.
If error remains don't use such a probe.
AcousticUnitType not supported by BF_Interface:
restart the system
Array bounds exceeded
restart the system
Attempt to read Probe-ID from an invalid probe connector.
reboot system.
AVI Save function fails
check the SCSI and the Power cable - restart the machine
and try again
B_Enhance Out Of Range
press ok and save this user-setting once again
B_Gain Out Of Range
press ok and save this user-setting once again
B_Reject Out Of Range
press ok and save this user-setting once again
B_TxFocus - not calculated and B_SHOT_PART_ON
restart the system
Backup error while writing. (Error during writing of backup data.)
check storage destination for Full Backup
(e.g., DVD/CD not empty, insufficient rights on target
Network drive, write protection on MO or USB-drive)
Backup error while verifying. (Checksum mismatch)
repeat backup
BC Ensemble Out Of Range
press ok and save this user-setting once again
BC lines_per_sequenz < 1
restart the system
BC lines_per_sequenz < 2
restart the system
BC_Dynamic Out Of Range
press ok and save this user-setting once again
BC_Gain Out Of Range
press ok and save this user-setting once again
BC_Lines: BC_LineDensity out of limit
restart the system
BCMC_Balance Out Of Range
press ok and save this user-setting once again
bCP_ConvertParameters failed
restart the system
bCP_ProcessIQEnsemblePacket failed
restart the system
bCP_ProcessIQMModePacket failed
restart the system