Section 1-3 - Safety Considerations
1-3-3
Mechanical Safety
(cont’d)
NOTE:
Special care should be taken when transporting the unit in a vehicle:
•
Secure the unit in an upright position.
•
Lock the wheels (brake)
•
DO NOT use the Control Panel as an anchor point.
•
Place the probes in their carrying case.
•
Eject any DVD, CD or Magneto Optical disk (MOD) from their drive.
1-3-4
Electrical Safety
1-3-4-1
Safe Practices
To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
The system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. If an extension cord is used with the system, make sure that the total
current rating of the system does not exceed the extension cord rating.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
1-3-4-2
Probes
Follow the guidelines before connecting a probe to the scanner:
•
Inspect the probe prior to each use for damage or degradation to the:
-
Housing
-
Cable strain relief
-
Lens
-
Seal
•
Do not use a damaged or defective probe.
•
Never immerse the probe connector or adapter into any liquid.
CAUTION
Keep the heat venting holes on the monitor unobstructed to avoid overheating of the monitor.
WARNING
WARNING
CONNECTING A VOLUSON® 730PRO / 730PROV SCANNER TO THE WRONG
VOLTAGE LEVEL WILL MOST LIKELY DESTROY THE SCANNER.